The Science Journal of the American Association for Respiratory Care

2002 OPEN FORUM Abstracts

Use of the Deltatrac II with the NPB 840 and NBP 7200 Ventilators

Allan Andrews, MS, RRT; Carl Haas, MLS, RRT; Ronald Dechert, MS, RRT; Critical Care Support Services, University of Michigan, Ann Arbor, MI

Background: Until recently resting energy expenditure (REE) calculations on ventilated patients were done using the Delta Trac II metabolic cart (SensorMedics Corporation, Yorba Linda, CA) in conjunction with the PB 7200 (Puritan Bennett, Inc., Carlsbad, CA), with acceptable results. The PB 840 is now our standard ventilator and it has been suggested that the 840 does not allow valid REE calculation. This study was done to compare the results obtained via the 7200 and those via the 840.

Methods: Ten normal subjects, breathing room air through a mouthpiece, had REE’s calculated under 5 conditions; no support, PS=5 cm H20 via the 7200 and 840, and with a VT=0.7 L & f=2 via 7200 and 840. The order of testing was randomized and done in duplicate

Results: The figure shows minimal difference between 7200 and 840 for pressure (8 of 10 subjects) or volume (7 of 10) breaths. There was no significant difference (Wilcoxon Signed Ranks Test) in REE’s for: 1) VC-840 vs. PS-840, 2) VC-7200 vs. PS-7200, 3) VC-840 vs. VC-7200, or $) PS-840 vs. PS-700.

Conclusions: Under these test conditions 1) There were no significant differences in REE’s between the 840 or 7200, for either breath type; 2) most REE’s were higher during mechanical ventilation. Due to the small sample size and limited variety of ventilator settings, further investigation is needed.

OF-02-143

You are here: RCJournal.com » Past OPEN FORUM Abstracts » 2002 Abstracts » Use of the Deltatrac II with the NPB 840 and NBP 7200 Ventilators