CORRELATION OF THE SENTEC DIGITAL MONITOR TRANSCUTANEOUS CO2 MEASUREMENT TO ARTERIAL CO2 USING ALTERNATE SITE PLACEMENT PROBES

Amber Chitwood, Aaron Light, Doug Pursley, Susan Howard, Ashley Ray, April Fritz, Alison Sorensen; Ozarks Technical Community College, Springfield, MO

Background: Previous studies have compared the TcCO2 measurement from the ear probe of the SenTec Digital Monitor (SenTec AG, Therwil, Switzerland) to PaCO2. In this study, we sought to evaluate the ability of SenTec’s V-Sign™ multi-site placement probe at monitoring TcCO2 using three different sites recommended by the manufacturer. Our hypothesis is that the three sites will accurately correlate with PaCO2. Methods: After obtaining IRB approval from our institution, we recruited 15 healthy adults already enrolled in one of two other institutional research projects. The other projects required participants to breathe oxygen from a particular oxygen delivery device while an arterial blood gas was performed. This study was performed simultaneously with one of the other two projects. After subjects were seated, V-Sign™ sensors were placed on the subject’s right clavicle, forehead above the right eyebrow, and left mastoid area. Participants were then instructed to relax, breathe normally, and not talk for a period of 15 minutes. During this period of time, stabilization of the SenTec Digital Monitor (SDM) was achieved. After a fifteen minute period, we performed a radial artery blood gas and measured PaCO2 using a GEM 3000 blood gas analyzer. No air bubbles were observed in any of the syringes and all samples were analyzed within five minutes. Upon flash of blood in the syringe, TcCO2 values were transcribed from the three SDMs. Between each individual, the sensors were cleaned and calibrated using recommended manufacturer’s guidelines. The SDM associated with each site placement was also rotated to ensure that any variance in values was not associated with a particular monitor. Results: All three site placements showed strong positive correlation with PaCO2. The forehead site showed no statistical difference between mean TcCO2 and mean PaCO2 using a paired t-test, whereas the mastoid and clavicle site’s mean readings were both 3.1 mmHg lower than actual PaCO2 and had a 0.57-1.0 mmHg higher standard deviation. Conclusion: All three multi-site placements that we evaluated offer clinicians reliable correlation between TcCO2 and PaCO2. Although the mastoid and clavicle sites showed statistical difference (p < 0.05) from PaCO2, the clinical difference appears minimal when one looks at the proximity of the means and standard deviations. Further research needs to be performed in the clinical setting to see if our results are similar in patients. Sponsored Research - None

You are here: RCJournal.com » Past OPEN FORUM Abstracts » 2009 Abstracts » CORRELATION OF THE SENTEC DIGITAL MONITOR TRANSCUTANEOUS CO2 MEASUREMENT TO ARTERIAL CO2 USING ALTERNATE SITE PLACEMENT PROBES