AARC Clinical Practice Guideline

Capnography/Capnometry during Mechanical Ventilation—2003 Revision & Update

CO2 MV 1.0 PROCEDURE:
Capnography comprises the continuous analysis and recording of carbon dioxide concentrations [CO2] in respiratory gases. Although the terms capnography and capnometry are sometimes considered synonymous, capnometry suggests measurement (ie, analysis alone) without a continuous
written record or waveform. Capnographic waveforms may be time-based or volume-based.

CO2 MV 2.0 DESCRIPTION/DEFINITION:
For the purposes of this Guideline, capnography refers to the evaluation of the [CO2] in the respiratory gases of mechanically ventilated patients. A capnographic device incorporates one of two types of analyzers: mainstream or sidestream.1 , 2 Mainstream analyzers insert a sampling window into the ventilator circuit for measurement of CO2, whereas a sidestream analyzer aspirates gas from the ventilator circuit, and the analysis occurs away from the ventilator circuit. Analyzers utilize infrared, mass or Raman spectra, or a photoacoustic spectra technology.3 - 5 Flow measuring devices are utilized in volume-based capnographs.

CO2 MV 3.0 SETTING:
This procedure may be performed by trained health care personnel in any setting in which mechanically ventilated patients are found—for example, the hospital, the extended care facility, or during transport.

CO2 MV 4.0 INDICATIONS:
On the basis of available evidence, capnography should not be mandated for all patients receiving mechanical ventilatory support, but it may be indicated for:

4.1 Evaluation of the exhaled [CO2], especially end-tidal CO2, which is the maximum partial pressure of CO2 exhaled during a tidal breath (just prior to the beginning of inspiration) and is designated PetCO2 6,7
4.2 Monitoring severity of pulmonary disease5 and evaluating response to therapy, especially therapy intended to improve the ratio of dead
space to tidal volume (VD/VT)8,9 and the matching of ventilation to perfusion (V/Q) ,10 and, possibly, to increase coronary blood flow6,11,12
4.3 Use as an adjunct to determine that tracheal rather than esophageal intubation has taken place (Low or absent cardiac output may negate its use for this indication.);13-18 colorimetric CO2 detectors are adequate devices for this purpose.19
4.4 Continued monitoring of the integrity of the ventilatory circuit, including the artificial airway13,16
4.5 Evaluation of the efficiency of mechanical ventilatory support by determination of the difference between the arterial partial pressure for CO2 (PaCO2) and the PetCO2 20,21
4.6 Monitoring adequacy of pulmonary, systemic, and coronary blood flow11,12,22-26
4.6.1 Estimation of effective (nonshunted) pulmonary capillary blood flow by a partial rebreathing method27-29
4.6.2 Use as an adjunctive tool to screen for pulmonary embolism (Evidence for the utility of dead space determinations as a screening tool for pulmonary embolism is at present not conclusive.)30,31
4.6.3 Monitoring the matching of ventilation to perfusion during independent lung ventilation for unilateral pulmonary contusion32
4.7 Monitoring inspired CO2 when CO2 gas is being therapeutically administered33
4.8 Graphic evaluation of the ventilator-patient interface; evaluation of the shape of the capnogram may be useful in detecting rebreathing of
CO2, obstructive pulmonary disease, waning neuromuscular blockade (‘curare cleft’), cardiogenic oscillations, esophageal intubation, cardiac arrest, and contamination of the monitor or sampling line with secretions or mucus.6,34
4.9 Measurement of the volume of CO2 elimination to assess metabolic rate and/or alveolar ventilation35

CO2 MV 5.0 CONTRAINDICATIONS:
There are no absolute contraindications to capnography in mechanically ventilated patients provided that the data obtained are evaluated with consideration given to the patient’s clinical condition.

CO2 MV 6.0 HAZARDS/COMPLICATIONS:
Capnography with a clinically approved device is a safe, noninvasive test, associated with few hazards. With mainstream analyzers, the use of too large a sampling window may introduce an excessive amount of dead space into the ventilator circuit.2,36 Care must be taken to minimize the amount of additional weight placed on the artificial airway by the addition of the sampling window or, in the case of a sidestream analyzer, the sampling line.

CO2 MV 7.0 LIMITATIONS OF PROCEDURE OR DEVICE:
Capnography, when performed using a device calibrated and operated as recommended by the manufacturer, has few limitations. It is important
to note that although the capnograph provides valuable information about the efficiency of ventilation (as well as pulmonary, systemic, and
coronary perfusion), it is not a replacement or substitute for assessing the PaCO2.20,34,37-41 The difference between PetCO2 and PaCO2 increases as dead-space volume increases. In fact, the difference between the PaCO2 and PetCO2 has been shown to vary within the same patient over time.42-47 Alterations in breathing pattern and tidal volume may introduce error into measurements designed to be made during stable, steady-state conditions.48-50 Interpretation of results must take into account the stability of physiologic parameters such as minute ventilation, tidal volume, cardiac output, ventilation/perfusion ratios and CO2 body stores.51,52 Certain situations may affect the reliability of the capnogram. The extent to which the reliability is affected varies somewhat among types of devices (infrared,53 photoacoustic, mass spectrometry, and Raman spectrometry). Limitations include:

7.1 The composition of the respiratory gas mixture may affect the capnogram (depending on the measurement technology incorporated).

7.1.1 The infrared spectrum of CO2 has some similarities to the spectra for both oxygen and nitrous oxide.53 High concentrations of either or both oxygen or nitrous oxide may affect the capnogram, and, therefore, a correction factor should be incorporated into the calibration of any capnograph used in such a setting.54
7.1.2 The reporting algorithm of some devices (primarily mass spectrometers) assumes that the only gases present in the sample are those that the device is capable of measuring.55 When a gas that the mass spectrometer cannot detect (such as helium) is present, the reported values of CO2 are incorrectly elevated in proportion to the concentration of helium present.

7.2 The breathing frequency may affect the capnograph. High breathing frequencies may exceed the response capabilities of the capnograph. 56 In addition, breathing frequency above 10 breaths/min has been shown to affect devices differently.57
7.3 The presence of Freon (used as a propellant in metered dose inhalers) in the respiratory gas has been shown to artificially increase the CO2 reading of mass spectrometers (ie, to show an apparent increase in [CO2]). A similar effect has not yet been demonstrated with Raman or infrared spectrometers.58
7.4 Contamination of the monitor or sampling system by secretions or condensate, a sample tube of excessive length, a sampling rate that is too high, or obstruction of the sampling chamber can lead to unreliable results.
7.5 Use of filters between the patient airway and the sampling line of the capnograph may lead to lowered PetCO2 readings.59,60
7.6 Low cardiac output may cause a false negative result when attempting to verify endotracheal tube (ETT) position in the trachea. False positive results have been reported with ETT position in the pharynx and when antacids and/or carbonated beverages are present in the stomach.18,61
7.7 Decreased tidal volume delivery is possible during volume modes, some dual control modes, and time-cycled pressure limited ventilation with low continuous flowrates if the sampling flowrate of a sidestream analyzer is too high, especially in neonates and pediatrics.
7.8 Inaccurate measurement of expired CO2 may be caused by leaks of gas from the patient/ ventilator system preventing collection of expired gases, including

7.8.1 Leaks in the ventilator circuit62
7.8.2 Leaks around tracheal tube cuffs or uncuffed tubes62

CO2 MV 8.0 ASSESSMENT OF NEED:
Capnography is considered a standard of care during anesthesia.1 3 , 6 3 The American Society of Anesthesiologists has suggested that capnography be available for patients with acute ventilatory failure on mechanical ventilatory support.63 The American College of Emergency Physicians recommends capnography as an adjunctive method to ensure proper ETT position.64 The international guidelines for emergency cardiovascular care recommend use of capnography to verify ETT placement in all age groups.65 Assessment of the need to use capnography with a specific patient should be guided by the clinical situation. The patient’s primary cause of respiratory failure and the acuteness of his or her condition should be considered.

CO2 MV 9.0 ASSESSMENT OF OUTCOME:
Results should reflect the patient’s condition and should validate the basis for ordering the monitoring. Documentation of results (along with all ventilatory and hemodynamic variables available), therapeutic interventions, and/or clinical decisions made based on the capnogram should be included in the patient’s chart.

CO2 MV 10.0 RESOURCES:
10.1 Equipment: The capnograph and accessories (eg, airway adapter, sampling tube, depending on capnograph). The capnograph should be calibrated as recommended by the manufacturer.
10.2 Personnel: Licensed or credentialed respiratory care practitioners or individuals with similar credentials (eg, MD, RN) who have the necessary training and demonstrated skills to correctly calibrate and evaluate the capnograph, assess the patient and the patient-ventilator system, and the ability to exercise appropriate clinical judgment.

CO2 MV 11.0 MONITORING:
During capnography the following should be considered and monitored:
11.1 Ventilatory variables: tidal volume, respiratory rate, positive end-expiratory pressure, inspiratory- to-expiratory time ratio (I:E), peak airway pressure, and concentrations of respiratory gas mixture36,37
11.2 Hemodynamic variables: systemic and pulmonary blood pressures, cardiac output, shunt, and ventilation-perfusion imbalances36,37

CO2 MV 12.0 FREQUENCY:
Capnography (or, at least, capnometry) should be available during endotracheal intubation.13,14,16,37,64 Capnography is not indicated for every mechanically ventilated patient; however, when it is used, the measurement period should be long enough to allow determination of the PaCO2- PetCO2 difference, note changes in PaCO2-PetCO2 difference as a result of therapy, and allow interpretation of observed trends.

CO2 MV 13.0 INFECTION CONTROL:
No specific precautions are necessary although Standard Precautions (as described by the Centers for Disease Control & Prevention)66 and precautions designed to limit the spread of tuberculosis67 should always be implemented during patientcare.

13.1 The sensor (the portion of the device contacting the patient’s airway) should be subjected to high-level disinfection between patients, according to the manufacturer’s recommendations.
13.2 The monitor (the portion not contacting the patient or the patient’s airway) should be cleaned as needed according to manufacturer’s recommendations.

Revised by Charles D McArthur RRT RPFT, Immanuel St Joseph’s Mayo Health System, Mankato, Minnesota. and approved by the 2003 CPG Steering Committee

Original Publication: Respir Care 1995;40(12):1321-1324.

 

REFERENCES

  1. Hess DR, Branson RD. Noninvasive respiratory monitoring equipment. In: Branson RD, Hess DR, Chatburn RL, editors. Respiratory care equipment. Philadelphia: Lippincott; 1994:184-216.
  2. Block FE Jr, McDonald JS. Sidestream versus mainstream carbon dioxide analyzers. J Clin Monit 1992;8(2):139-141.
  3. O’Flaherty D. Capnometry. London: BMJ Publishing Group; 1994:21-54.
  4. VanWagenen RA, Westenskow DR, Benner RE, Gregonis DE, Coleman DL. Dedicated monitoring of anesthetic and respiratory gases by Raman scattering. J Clin Monit 1986;2(4):215-222.
  5. Cinnella G, Dambrosio M, Brienza N, Giuliani R, Bruno F, Fiore T, Brienza A. Independent lung ventilation in patients with unilateral pulmonary contusion: monitoring with compliance and etCO2. Intensive Care Med 2001;27(12):1860-1867.
  6. Carlon GC, Ray C Jr, Miodownik S, Kopec I, Groeger JS. Capnography in mechanically ventilated patients. Crit Care Med 1988;16(5):550-556.
  7. Gravenstein N, Good ML. Noninvasive assessment of cardiopulmonary function. In: Civetta JM, Taylor RW, Kirby RR, editors. Critical care. Philadelphia: Lippincott; 1988:291-311.
  8. Yamanaka MK, Sue DY. Comparison of arterial-endtidal PCO2 difference and dead space/tidal volume ratio in respiratory failure. Chest 1987;92(5):832-835.
  9. Poppius H, Korhonen O, Viljanen AA, Kreus KE. Arterial to end-tidal CO2 difference in respiratory disease. Scand J Respir Dis 1975;56(5):254-262.
  10. Burrows FA. Physiologic dead space, venous admixture, and the arterial to end-tidal carbon dioxide difference in infants and children undergoing cardiac surgery. Anesthesiology 1989;70(2):219-225.
  11. Kafer ER. Respiratory function in pulmonary thromboembolic disease. Am J Med 1969;47(6):904-915.
  12. Sanders AB, Atlas M, Ewy GA, Kern KB, Bragg S. Expired PCO2 as an index of coronary perfusion pressure. Am J Emerg Med 1985;3(2):147-149.
  13. Eichhorn JH, Cooper JB, Cullen DJ, Maier WR, Philip JH, Seeman RG. Standards for patient monitoring during anesthesia at Harvard Medical School. JAMA 1986;256(8):1017-1020.
  14. Birmingham PK, Cheney FW, Ward RJ. Esophageal intubation: a review of detection techniques. Anesth Analg 1986;65(8):886-891.
  15. Sanders AB. Capnometry in emergency medicine. Ann Emerg Med 1989;18(12):1287-1290.
  16. Murray IP, Modell JH. Early detection of endotracheal tube accidents by monitoring carbon dioxide concentration in respiratory gas. Anesthesiology 1983;59(4): 344-346.
  17. Roberts WA, Maniscalco WM, Cohen AR, Litman RS, Chhibber A. The use of capnography for recognition of esophageal intubation in the neonatal intensive care unit. Pediatr Pulmonol 1995;19(5):262-268.
  18. Li J. Capnography alone is imperfect for endotracheal tube placement confirmation during emergency intubation. J Emerg Med 2001;20(3):223-229.
  19. Goldberg JS. Rawle PR. Zehnder JL. Sladen RN. Colorimetric end-tidal carbon dioxide monitoring for tracheal intubation. Anesth Analg 1990;70(2):191-194.
  20. Graybeal JM, Russell GB. Capnometry in the surgical ICU: an analysis of the arterial-to-end-tidal carbon dioxide difference. Respir Care 1993;38(8):923-928.
  21. Braman SS, Dunn SM, Amico CA, Millman RP. Complications of intrahospital transport of critically ill patients. Ann Internal Med 1987;107(4):469-473.
  22. Kalenda Z. The capnogram as a guide to the efficacy of cardiac massage. Resuscitation 1978;6(4):259-263.
  23. Sanders AB, Ewy GA, Bragg S, Atlas M, Kern KB. Expired PCO2 as prognostic indicator of successful resuscitation from cardiac arrest. Ann Emerg Med 1985;14(10):948-952.
  24. Shibutani K, Muraoka M, Shirasaki S, Kubal K, Sanchala VT, Gupte P. Do changes in end-tidal PCO2 quantitatively reflect changes in cardiac output? Anesth Analg 1994;79(5):829-833.
  25. Weil MH, Bisera J, Trevino RP, Rackow EC. Cardiac output and end-tidal carbon dioxide. Crit Care Med 1985;13(11):907-909.
  26. Levine RL, Wayne MA, Miller CC. End-tidal carbon dioxide and outcome of out-of-hospital cardiac arrest. N Engl J Med 1997;337(5):301-306.
  27. de Abreu MG, Quintel M, Ragaller M, Albrecht DM. Partial carbon dioxide rebreathing: a reliable technique for noninvasive measurement of nonshunted pulmonary capillary blood flow. Crit Care Med 1997;25(4): 675-683.
  28. van Heerden PV, Baker S, Lim SI, Weidman C, Bulsara M. Clinical evaluation of the non-invasive cardiac output (NICO) monitor in the intensive care unit. Anaesth Intensive Care 2000;28(4):427-430.
  29. de Abreu MG, Geiger S, Winkler T, Ragaller M, Pfeiffer T, Leutheuser D, Albrecht DM. Evaluation of a new device for noninvasive measurement of nonshunted pulmonary capillary blood flow in patients with acute lung injury. Intensive Care Med 2002;28(3):318-323.
  30. Rodger MA, Jones G, Rasuli P, Raymond F, Djunaedi H, Bredeson CN, Wells PS. Steady-state end-tidal alveolar dead space fraction and D-dimer: bedside tests to exclude pulmonary embolism. Chest 2001;120(1): 115-119.
  31. Kline JA, Johns KL, Colucciello SA, Israel EG. New diagnostic tests for pulmonary embolism. Ann Emerg Med 2000;35(2):168-180.
  32. Colman Y, Krauss B. Microstream capnograpy technology: a new approach to an old problem. J Clin Monit Comput 1999;15(6):403-409.
  33. Fatigante L, Cartei F, Ducci F, Marini C, Predilletto R, Caciagli P, Laddaga M. Carbogen breathing in patients with gliblastoma multiforme submitted to radiotherapy: assessment of gas exchange parameters. Acta Oncol 1994;33(7):807-811.
  34. Bhavani-Shankar K, Moseley H, Kumar AY, Delphi Y. Capnometry and anesthesia. Can J Anesth 1992;39(6):617-632.
  35. Taskar V, John J, Larsson A, Wetterberg T, Jonson B. Dynamics of carbon dioxide elimination following ventilator resetting. Chest 1995;108(1):196-202.
  36. Szaflarski NL, Cohen NH. Use of capnography in critically ill adults. Heart Lung 1991;20(4):363-374. Erratum 1991;20(6):630
  37. Hess D. Capnometry and capnography: technical aspects, physiologic aspects, and clinical applications. Respir Care 1990;35(6):557-573.
  38. Isert P. Control of carbon dioxide levels during neuroanesthesia: current practice and an appraisal of our reliance upon capnography. Anaesth Intensive Care 1994;22(4):435-441.
  39. Jellinek H, Hiesmayr M, Simon P, Klepetko W, Haider W. Arterial to end-tidal CO2 tension difference after bilateral lung transplantation. Crit Care Med 1993;21(7):1035-1040.
  40. Laffon M, Gouchet A, Sitbon P, Guicheteau V, Biyick E, Duchalais A, Mercier C. Difference between arterial and end-tidal carbon dioxide pressures during laparoscopy in paediatric patients. Can J Anaesth 1998; 45(6):561-563.
  41. Kerr ME, Zempsky J, Sereika S, Orndoff P, Rudy EB. Relationship between arterial carbon dioxide and endtidal carbon dioxide in mechanically ventilated adults with severe head trauma. Crit Care Med 1996;24(5):785-790.
  42. Russell GB, Graybeal JM. End-tidal carbon dioxide as an indicator of arterial carbon dioxide in neurointensive care patients. J Neurosurg Anesth 1992;4:245-249.
  43. Russell GB, Graybeal JM. Reliability of the arterial to end-tidal carbon dioxide gradient in mechanically ventilated patients with multisystem trauma. J Trauma 1994;36(3):317-322.
  44. Russell GB, Graybeal JM. The arterial to end-tidal carbon dioxide difference in neurosurgical patients during craniotomy. Anesth Analg 1995;81(4):806-810.
  45. Hess D, Schlottag A, Levin B, Mathai J, Rexrode WO. An evaluation of the usefulness of end-tidal PC O2 to aid weaning from mechanical ventilation following cardiac surgery. Respir Care 1991;36(8):837-843.
  46. Grenier B, Verchere E, Mesli A, Dubreuil M, Siao D, Vandendriessche M, et al. Capnography monitoring during neurosurgery: reliability in relation to various intraoperative positions. Anesth Analg 1999;88(1): 43-48.
  47. Seguin P, Bleichner JP, Branger B, Guillou YM, Feuillu A, Malledant Y. The measurement of end-tidal carbon dioxide (PetCO2) is not a significant parameter to monitor in patients with severe traumatic brain injury. Can J Anaesth 2001;48(4):396-400.
  48. Wessel HU, Stout RL, Bastanier CK, Paul MH. Breathby- breath variation of FRC: effect on VO2 and VCO2 measured at the mouth. J Appl Physiol 1979;46(6):1122-1126.
  49. Gamma de Abreu M, Melo MF, Giannella-Neto A. Pulmonary capillary blood flow by partial CO2 rebreathing: importance of the regularity of the respiratory pattern. Clin Physiol 2000;20(5):388-398.
  50. Brandi LS, Bertolini R, Santini L, Cavani S. Effects of ventilator resetting on indirect calorimetry measurement in the critically ill surgical patient. Crit Care Med 1999;27(3):531-539.
  51. Farhi LE, Rahn H. Dynamics change in carbon dioxide stores. Anesthesiology 1960;21:604-614.
  52. Damask MC, Askanazi J, Weissman C, Elwyn DH, Kinney JM. Artifacts in measurement of resting energy expenditure. Crit Care Med 1983;11(9):750-752.
  53. Ammann ECB, Galvin RD. Problems associated with the determination of carbon dioxide by infrared absorption. J Appl Physiol 1968;25(3):333-335.
  54. Severinghaus JW, Larson CP, Eger EI. Correction factors for infrared carbon dioxide pressure broadening by nitrogen, nitrous oxide and cyclopropane. Anesthesiology 1961;22:429-432.
  55. Perkin-Elmer MGA 1100 Operation and maintenance manual. 1982:6-26.
  56. From RP, Scamman FL. Ventilatory frequency influences accuracy of end-tidal CO2 measurements: analysis of seven capnometers. Anesth Analg 1988;67(9): 884-886.
  57. Graybeal JM, Russell GB. Relative agreement between Raman and mass spectrometry for measuring end-tidal carbon dioxide. Respir Care 1994;39(3):190-194.
  58. Elliott WR, Raemer DB, Goldman DB, Philip JH. The effects of bronchodilator-inhaler aerosol propellants on respiratory gas monitors. J Clin Monit 1991;7(2): 175-180.
  59. Hardman JG, Curran J, Mahajan RP. End-tidal carbon dioxide measurement and breathing system filters. Anaesthesia 1997;52(7):646-648.
  60. Hardman JG, Mahajan RP, Curran J. The influence of breathing system filters on paediatric capnography. Paediatr Anaesth 1999;9(1):35-38.
  61. Sum Ping ST, Mehta MP, Symreng T. Reliability of capnography in identifying esophageal intubation with carbonated beverage or antacid in the stomach. Anesth Analg 1991;73(3):333-337.
  62. Branson RD. The measurement of energy expenditure: instrumentation, practical considerations, and clinical aplication. Respir Care 1990;35(7):640-656; discussion 656-659.
  63. American Society of Anesthesiologists: Standards for basic anesthetic monitoring. 1998. Available at: http//www.asahq.org/
  64. American College of Emergency Physicians. Clinical Policies Committee.Verification of endotracheal tube placement. Ann Emerg Med 2002;40(5):551-552.
  65. Guidelines 2000 for Cardiopulmonary resuscitation and emergency cardiovascular care: international consensus on science. Circulation 2000;102(8):I1-I370.
  66. Bolyard EA, Tablan OC, Williams WW, Pearson ML, Shapiro CN, Deitchmann SD. Guideline for infection control in healthcare personnel, 1998. Hospital Infection Control Practices Advisory Committee. Infect Control Hosp Epidemiol 1998;19(6):4407-4462. Erratum in: Infect Control Hosp Epidemiol 1998;19(7):4493
  67. Guidelines for preventing the transmission of Mycobacterium tuberculosis in health-care facilities, 1994. Centers for Disease Control and Prevention. MMWR Recomm Rep 1994;43(RR-13):1-132.

 

Interested persons may copy these Guidelines for noncommercial purposes of scientific or educational advancement. Please credit AARC and Respiratory Care Journal.