CO2 MV 1.0 PROCEDURE:
Capnography comprises the continuous analysis and recording of carbon
dioxide concentrations [CO2] in respiratory gases. Although the terms
capnography and capnometry are sometimes considered synonymous, capnometry suggests
measurement (ie, analysis alone) without a continuous
written record or waveform. Capnographic waveforms may be time-based or volume-based.
CO2 MV 2.0 DESCRIPTION/DEFINITION:
For the purposes of this Guideline, capnography refers to the evaluation of
the [CO2] in the respiratory gases of mechanically ventilated patients.
A capnographic device incorporates one of two types of analyzers: mainstream
or sidestream.1 , 2 Mainstream analyzers insert a sampling window
into the ventilator circuit for measurement of CO2, whereas a sidestream
analyzer aspirates gas from the ventilator circuit, and the analysis occurs
away from the ventilator circuit. Analyzers utilize infrared, mass or Raman
spectra, or a photoacoustic spectra technology.3 - 5 Flow measuring
devices are utilized in volume-based capnographs.
CO2 MV 3.0 SETTING:
This procedure may be performed by trained health care personnel in
any setting in which mechanically ventilated patients are found—for example,
the hospital, the extended care facility, or during transport.
CO2 MV 4.0 INDICATIONS:
On the basis of available evidence, capnography should not be mandated
for all patients receiving mechanical ventilatory support, but it may be indicated
for:
4.1 Evaluation of the exhaled [CO2], especially
end-tidal CO2, which is the maximum partial pressure of CO2
exhaled during a tidal breath (just prior to the beginning of inspiration) and
is designated PetCO2 6,7
4.2 Monitoring severity of pulmonary disease5 and
evaluating response to therapy, especially therapy intended to improve the ratio
of dead
space to tidal volume (VD/VT)8,9 and the matching
of ventilation to perfusion (V/Q) ,10 and, possibly, to increase
coronary blood flow6,11,12
4.3 Use as an adjunct to determine that tracheal rather
than esophageal intubation has taken place (Low or absent cardiac output may
negate its use for this indication.);13-18 colorimetric CO2
detectors are adequate devices for this purpose.19
4.4 Continued monitoring of the integrity of the ventilatory
circuit, including the artificial airway13,16
4.5 Evaluation of the efficiency of mechanical ventilatory
support by determination of the difference between the arterial partial pressure
for CO2 (PaCO2) and the PetCO2 20,21
4.6 Monitoring adequacy of pulmonary, systemic, and coronary
blood flow11,12,22-26
4.6.1 Estimation of effective (nonshunted) pulmonary capillary
blood flow by a partial rebreathing method27-29
4.6.2 Use as an adjunctive tool to screen for pulmonary embolism
(Evidence for the utility of dead space determinations as a screening tool for
pulmonary embolism is at present not conclusive.)30,31
4.6.3 Monitoring the matching of ventilation to perfusion during
independent lung ventilation for unilateral pulmonary contusion32
4.7 Monitoring inspired CO2 when CO2
gas is being therapeutically administered33
4.8 Graphic evaluation of the ventilator-patient interface;
evaluation of the shape of the capnogram may be useful in detecting rebreathing
of
CO2, obstructive pulmonary disease, waning neuromuscular blockade
(‘curare cleft’), cardiogenic oscillations, esophageal intubation,
cardiac arrest, and contamination of the monitor or sampling line with secretions
or mucus.6,34
4.9 Measurement of the volume of CO2 elimination
to assess metabolic rate and/or alveolar ventilation35
CO2 MV 5.0 CONTRAINDICATIONS:
There are no absolute contraindications to capnography in mechanically ventilated
patients provided that the data obtained are evaluated with consideration given
to the patient’s clinical condition.
CO2 MV 6.0 HAZARDS/COMPLICATIONS:
Capnography with a clinically approved device is a safe, noninvasive
test, associated with few hazards. With mainstream analyzers, the use of too
large a sampling window may introduce an excessive amount of dead space into
the ventilator circuit.2,36 Care must be taken to minimize the amount
of additional weight placed on the artificial airway by the addition of the
sampling window or, in the case of a sidestream analyzer, the sampling line.
CO2 MV 7.0 LIMITATIONS OF PROCEDURE OR DEVICE:
Capnography, when performed using a device calibrated and operated as recommended
by the manufacturer, has few limitations. It is important
to note that although the capnograph provides valuable information about the
efficiency of ventilation (as well as pulmonary, systemic, and
coronary perfusion), it is not a replacement or substitute for assessing the
PaCO2.20,34,37-41 The difference between PetCO2
and PaCO2 increases as dead-space volume increases. In fact, the
difference between the PaCO2 and PetCO2 has been shown
to vary within the same patient over time.42-47 Alterations in breathing
pattern and tidal volume may introduce error into measurements designed to be
made during stable, steady-state conditions.48-50 Interpretation
of results must take into account the stability of physiologic parameters such
as minute ventilation, tidal volume, cardiac output, ventilation/perfusion ratios
and CO2 body stores.51,52 Certain situations may affect
the reliability of the capnogram. The extent to which the reliability is affected
varies somewhat among types of devices (infrared,53 photoacoustic,
mass spectrometry, and Raman spectrometry). Limitations include:
7.1 The composition of the respiratory gas mixture may affect the capnogram (depending on the measurement technology incorporated).
7.1.1 The infrared spectrum of CO2 has some similarities to the spectra for both oxygen and nitrous oxide.53 High concentrations of either or both oxygen or nitrous oxide may affect the capnogram, and, therefore, a correction factor should be incorporated into the calibration of any capnograph used in such a setting.54
7.1.2 The reporting algorithm of some devices (primarily mass spectrometers) assumes that the only gases present in the sample are those that the device is capable of measuring.55 When a gas that the mass spectrometer cannot detect (such as helium) is present, the reported values of CO2 are incorrectly elevated in proportion to the concentration of helium present.7.2 The breathing frequency may affect the capnograph. High breathing frequencies may exceed the response capabilities of the capnograph. 56 In addition, breathing frequency above 10 breaths/min has been shown to affect devices differently.57
7.3 The presence of Freon (used as a propellant in metered dose inhalers) in the respiratory gas has been shown to artificially increase the CO2 reading of mass spectrometers (ie, to show an apparent increase in [CO2]). A similar effect has not yet been demonstrated with Raman or infrared spectrometers.58
7.4 Contamination of the monitor or sampling system by secretions or condensate, a sample tube of excessive length, a sampling rate that is too high, or obstruction of the sampling chamber can lead to unreliable results.
7.5 Use of filters between the patient airway and the sampling line of the capnograph may lead to lowered PetCO2 readings.59,60
7.6 Low cardiac output may cause a false negative result when attempting to verify endotracheal tube (ETT) position in the trachea. False positive results have been reported with ETT position in the pharynx and when antacids and/or carbonated beverages are present in the stomach.18,61
7.7 Decreased tidal volume delivery is possible during volume modes, some dual control modes, and time-cycled pressure limited ventilation with low continuous flowrates if the sampling flowrate of a sidestream analyzer is too high, especially in neonates and pediatrics.
7.8 Inaccurate measurement of expired CO2 may be caused by leaks of gas from the patient/ ventilator system preventing collection of expired gases, including7.8.1 Leaks in the ventilator circuit62
7.8.2 Leaks around tracheal tube cuffs or uncuffed tubes62
CO2 MV 8.0 ASSESSMENT OF NEED:
Capnography is considered a standard of care during anesthesia.1 3 , 6 3 The
American Society of Anesthesiologists has suggested that capnography be available
for patients with acute ventilatory failure on mechanical ventilatory support.63
The American College of Emergency Physicians recommends capnography as an adjunctive
method to ensure proper ETT position.64 The international guidelines
for emergency cardiovascular care recommend use of capnography to verify ETT
placement in all age groups.65 Assessment of the need to use capnography
with a specific patient should be guided by the clinical situation. The patient’s
primary cause of respiratory failure and the acuteness of his or her condition
should be considered.
CO2 MV 9.0 ASSESSMENT OF OUTCOME:
Results should reflect the patient’s condition and should validate the
basis for ordering the monitoring. Documentation of results (along with all
ventilatory and hemodynamic variables available), therapeutic interventions,
and/or clinical decisions made based on the capnogram should be included in
the patient’s chart.
CO2 MV 10.0 RESOURCES:
10.1 Equipment: The capnograph and accessories (eg, airway
adapter, sampling tube, depending on capnograph). The capnograph should be calibrated
as recommended by the manufacturer.
10.2 Personnel: Licensed or credentialed respiratory care practitioners
or individuals with similar credentials (eg, MD, RN) who have the necessary
training and demonstrated skills to correctly calibrate and evaluate the capnograph,
assess the patient and the patient-ventilator system, and the ability to exercise
appropriate clinical judgment.
CO2 MV 11.0 MONITORING:
During capnography the following should be considered and monitored:
11.1 Ventilatory variables: tidal volume, respiratory rate,
positive end-expiratory pressure, inspiratory- to-expiratory time ratio (I:E),
peak airway pressure, and concentrations of respiratory gas mixture36,37
11.2 Hemodynamic variables: systemic and pulmonary blood pressures,
cardiac output, shunt, and ventilation-perfusion imbalances36,37
CO2 MV 12.0 FREQUENCY:
Capnography (or, at least, capnometry) should be available during endotracheal
intubation.13,14,16,37,64 Capnography is not indicated for every
mechanically ventilated patient; however, when it is used, the measurement period
should be long enough to allow determination of the PaCO2- PetCO2
difference, note changes in PaCO2-PetCO2 difference as
a result of therapy, and allow interpretation of observed trends.
CO2 MV 13.0 INFECTION CONTROL:
No specific precautions are necessary although Standard Precautions (as described
by the Centers for Disease Control & Prevention)66 and precautions
designed to limit the spread of tuberculosis67 should always be implemented
during patientcare.
13.1 The sensor (the portion of the device contacting the
patient’s airway) should be subjected to high-level disinfection between
patients, according to the manufacturer’s recommendations.
13.2 The monitor (the portion not contacting the patient or
the patient’s airway) should be cleaned as needed according to manufacturer’s
recommendations.
Revised by Charles D McArthur RRT RPFT, Immanuel St Joseph’s Mayo Health System, Mankato, Minnesota. and approved by the 2003 CPG Steering Committee
Original Publication: Respir Care 1995;40(12):1321-1324.