You are here: RCJournal.com » Clinical Practice Guidelines
Reprinted from the September 2003 issue of RESPIRATORY CARE [Respir Care 2003;48(9):869–879]
Summary of Recommendations
Introduction
A concern related to the care of the mechanically ventilated patient is the
development of VAP. For many years this concern focused on the ventilator
circuit and humidifier. Accordingly, the circuit and humidifier have been
changed on a regular basis in an attempt to decrease the VAP rate. However,
as the evidence evolved, it became apparent that the origin of VAP is more
likely from sites other than the ventilator circuit,1,2 and thus
the prevailing practice has become one of changing circuits less frequently.
3 If this practice is safe, it will offer substantial cost savings.
Other issues related to the components of the circuit and VAP have also become
more important recently. For example, humidification systems can be either
active or passive. Increasingly, inline suction is used, and this becomes
part of the ventilator circuit.
A systematic review of the literature was conducted with the intention of making recommendations for change frequency of the ventilator circuit and additional components of the circuit. Specifically, the Writing Committee wrote these evidence-based clinical practice guidelines to address the following questions:
1. Do ventilator circuits need to be changed at regular intervals:
2. What is the economic impact of decreasing the frequency of ventilator circuit changes?
3. What are the issues related to circuit type?4. Does the choice of active versus passive humidification affect ventilator circuit change frequency?
5. Do passive humidifiers need to be changed at regular intervals:
6. Do in-line suction catheters need to be changed at regular intervals:
7. Are there specific populations for which the recommendations should be altered?
Methods
To identify the evidence for addressing these questions, a PubMed (MEDLINE)
search was conducted using the following search terms: pneumonia AND mechanical
ventilation, humidifier, ventilator circuit, heated circuit, suction catheter,
endotracheal suction, closed suction catheter, respiratory therapy equipment,
endotracheal intubation, heat and moisture exchanger, tracheostomy, respiratory
care, equipment contamination, equipment disinfection, artificial ventilation.
The search was confined to human studies published in the English language.
References and abstracts were retrieved into reference management software
(EndNote, ISI, Berkeley, California). By inspection of these titles, references
having no relevance to the study questions were eliminated. For the titles
that remained, the abstracts were assessed for relevance and additional references
were eliminated as appropriate. This process was conducted independently by
2 individuals, after which their reference lists were merged to provide the
reference base for further analysis. Throughout the process of developing
these guidelines, members of the Writing Committee surveyed cross-references
to identify additional references to be added to the reference base for analysis.
Data were extracted from selected references using a standardized critique form. To validate this form and to establish the reliability of the review process, several references were evaluated by the entire committee during a face-to-face meeting. All references were then independently examined by at least 2 members of the Writing Committee. The critiques were compared and differences were resolved using an iterative process. All references were graded according to the following scheme:
Level 1: Randomized, controlled trial with statistically significant results
Level 2: Randomized, controlled trial with significant threats to validity (eg, small sample size, inappropriate blinding, weak methodology)
Level 3: Observational study with a concurrent control group
Level 4: Observational study with a historical control group
Level 5: Bench study, animal study, case series
The critique forms were submitted to the principal author of the guideline (DRH), who transferred the information into evidence tables and conducted appropriate statistical analysis.
Quantitative analysis consisted of meta-analysis and petograms. Statistical analysis was conducted using RevMan software (RevMan Analyses, Version 1.0 for Windows, in Review Manager [RevMan] 4.2, Oxford, England: The Cochrane Collaboration, 2003). Relative risk was calculated using a random effect model. P = 0.05 was considered statistically significant. Following a systematic review of the literature, recommendations were drafted by the Writing Committee and assigned one of the following grades, based on the strength of the evidence:
Grade A: Scientific evidence provided by randomized, well-designed, well-conducted, controlled trials with statistically significant results that consistently support the guideline recommendation; supported by Level 1 or 2 evidence
Grade B: Scientific evidence provided by well-designed, well-conducted observational studies with statistically significant results that consistently support the guideline recommendation; supported by Level 3 or 4 evidence
Grade C: Scientific evidence from bench studies, animal studies, case studies; supported by Level 5 evidence
Grade D: Expert opinion provides the basis for the guideline recommendation, but scientific evidence either provided inconsistent results or was lacking
The draft document was then reviewed by experts on ventilator circuit care. Each of the reviewer’s comments was carefully assessed and the document was further revised as appropriate.
Do Ventilator Circuits Need to Be Changed at Regular Intervals?
Based on studies published in the 1960s that showed an association between respiratory equipment and nosocomial pneumonia,4,5 the practice of changing ventilator circuits at least daily was established. In fact, circuits were changed every 8 hours in some hospitals, in an attempt to reduce the incidence of VAP. This practice was challenged in a landmark study published by Craven et al in 1982.6 In that study 240 cultures of inspiratory-phase gas were obtained from 95 patients. There was no significant difference in the frequency of positive cultures in circuits changed every 24 hours (30%) and in circuits changed every 48 hours (32%). Moreover, no significant increase in circuit colonization occurred between 24 and 48 hours. Based on that study, most hospitals in the United States adopted the practice of changing ventilator circuits at 48-hour intervals. Craven et al estimated that $300,000 (in 1982 dollars) would be saved at 20 Boston teaching hospitals by adopting this practice. Interestingly, Craven et al did not report VAP rates in their study.
Fig. 1. Randomized, controlled trials of the
relationship between ventilator circuit change frequency and
the risk of ventilator-associated pneumonia. RR = relative risk. CI = confidence
interval.

The effect of ventilator circuit change interval on VAP rate was assessed in 4 prospective randomized, controlled trials (Table 1 and Fig. 1).7–10 Although each of these studies evaluated different circuit change intervals, the combined effect supports the practice of less frequent circuit changes (relative risk 0.76, 95% confidence interval [CI] 0.57 to 1.00, p = 0.05). In each of these studies, the risk of VAP was decreased when circuits were changed less frequently. Ventilator circuit change interval was also assessed in 7 studies with historical control groups (Table 2 and Fig. 2).11–17 Again, the combined effect supports the practice of less frequent circuit changes (relative risk 0.87, 95% CI 0.63 to 1.18, p = 0.37). Two well-designed randomized, controlled trials evaluated the practice of “no changes” of ventilator circuits.8,9 However, the maximum duration of time that the circuit can be used safely is unknown. In one of those studies, the maximum duration of use of a circuit was 29 days.8 The other study did not report the maximum duration of use of a circuit but did report that 35% of patients were ventilated for > 14 days.9
The costs associated with ventilator circuit changes were calculated in 8 studies.6,8,9,12–15,17 Because these studies were conducted over a span of 20 years and in different countries, direct cost comparisons are difficult. Not surprisingly, each of these studies suggests considerable savings in personnel and materials costs with less frequent ventilator circuit changes. One study evaluated equipment failure (circuit leaks) related to ventilator circuit change frequency.9 In that study there was no significant difference in equipment failures when circuits were changed at weekly intervals and when circuits were not changed at regular intervals.
Table 1. Summary of Randomized Controlled
Trials Investigating the Relationship
Between Ventilator Circuit Change Frequency and the Risk of Ventilator-Associated
Pneumonia
| Citation |
Study
Population |
Blinding |
VAP
Diagnosis |
Control
Group |
Treatment
Group |
Control Group |
Treatment Group |
Level |
Relative
Risk (95% CI) |
||
| n |
Pneumonia
(%) |
n |
Pneumonia
(%) |
||||||||
Craven 19867 |
Adult patients requiring mechanical ventilation > 48 h |
VAP assessors |
Clinical |
Circuit changes every 24 h |
Circuit changes every 48 h |
106 |
29.2 |
127 |
14.2 |
1 |
0.48 (0.29, 0.82) |
Dreyfuss 19918 |
Adult patients requiring mechanical ventilation > 48 h |
VAP assessors |
Quantitative cultures |
Circuit changes every 48 h |
No circuit changes |
35 |
31.4 |
28 |
28.5 |
1 |
0.91 (0.42, 1.95) |
Kollef 19959 |
Adult patients requiring mechanical ventilation > 5 d |
VAP assessors |
Clinical |
Circuit changes every 7 d |
No circuit changes |
153 |
28.8 |
147 |
24.5 |
1 |
0.85 (0.58, 1.24) |
Long 199610 |
Neonatal and adult mechanically ventilated patients |
None |
Clinical |
Circuit changes 3 times/wk |
Circuit change 1/wk |
213 |
12.7 |
234 |
11.1 |
2 |
0.88 (0.53, 1.45) |
TOTAL |
507 |
22.3 |
536 |
16.4 |
0.76 (0.57, 1.00) |
||||||
VAP = ventilator-associated
pneumonia. CI = confidence interval. |
|||||||||||
Fig. 2. Observational studies of the relationship
between ventilator circuit change
frequency and the risk of ventilator-associated pneumonia. RR = relative risk.
CI = confidence interval.

The majority of the studies were in adult patients in acute care units. One study was conducted in a subacute care unit.13 Two studies included neonatal and pediatric mechanically ventilated patients.10,11 Although ventilator circuit change interval has been studied less in patient groups other than adult patients in acute care units, the available evidence suggests no increased risk for VAP associated with infrequent circuit changes in these populations. There have been no studies that separately addressed special populations such as immunocompromised or burned patients.
Recommendation #1. Ventilator circuits should not be changed routinely for infection control purposes. The available evidence suggests no patient harm and considerable cost savings associated with extended ventilator circuit change intervals. The maximum duration of time that circuits can be used safely is unknown. (Grade A)
Most studies used heated passover humidifiers, although several used bursting-bubble cascade-type humidifiers. 8,12,17 There is concern related to the use of bursting-bubble humidifiers because these have shown the potential to generate aerosols capable of carrying microorganisms.18,19 However, this is not a consideration in the present day, as these devices are no longer commercially available. Moreover, bacterial levels in heated humidifiers are low and nosocomial pathogens survive poorly in this environment.20 Although a few studies used reusable circuits,11,17 most used disposable circuits. In several studies, heated-wire circuits were used.10,14 –16 One small study21 compared heated-wire circuits and nonheated-wire circuits and found no difference in VAP rate (relative risk 1.57 in favor of nonheated- wire circuits, 95% CI 0.55 to 4.45). The condensate that accumulates in the ventilator circuit is contaminated,22 and care should be taken to avoid its cross-contamination of other patients. Although it makes sense that care should be taken to avoid breaking the circuit—which could contaminate the interior of the ventilator circuit—this has not been studied. One observational study compared daily with biweekly circuit changes with the use of a passive humidifier and reported no change in VAP rate with the longer circuit change interval.23 Another study compared disposable and reusable humidifiers during mechanical ventilation, and reported no difference in VAP.24 It is fair to say that the risk of VAP is not increased by less frequent circuit changes, despite a variety of practices related to the type of circuit used and the care of the circuit. Standard practice calls for use of sterile water in the humidifier of the ventilator circuit. Because water is an important reservoir for nosocomial pathogens and there has been no published study of this topic using modern humidification systems, the practice of filling humidifiers with sterile water appears appropriate.
Table 2. Summary of Observational Studies
Investigating the Relationship Between
Ventilator Circuit Change Frequency and the Risk of Ventilator-Associated
Pneumonia
| Citation |
Study
Population |
VAP
Diagnosis |
Control
Group |
Treatment
Group |
Control Group |
Treatment Group |
Level |
Relative
Risk (95% CI) |
||
| n |
Pneumonia
(%) |
n |
Pneumonia
(%) |
|||||||
Lareau 197811 |
Adult, pediatric, and neonatal mechanically
ventilated patients |
Clinical |
Circuit changes at 8-h intervals |
Circuit changes at 24-h intervals |
213 |
7.5 |
271 |
11.8 |
4 |
1.57 (0.89, 2.79) |
Hess 199512 |
Adult mechanically ventilated patients
|
Clinical |
Circuit changes at 2-d intervals
|
Circuit changes at 7-d intervals
|
1,708 |
5.6 |
1,715 |
4.6 |
4 |
0.83 (0.62, 1.11) |
Thompson 199613 |
Adult mechanically ventilated patients
in a subacute care facility |
Clinical |
Circuit changes at 7-d intervals
|
Circuit changes at 14-d intervals
|
31 |
9.7 |
18 |
11.1 |
4 |
1.15 (0.21, 6.24) |
Kotilainen 199714 |
Adult mechanically ventilated patients
|
Clinical |
Circuit changes at 3-d intervals
|
Circuit changes at 7-d intervals
|
88 |
9.1 |
146 |
6.2 |
4 |
0.68 (0.27, 1.69) |
Fink 199815 |
Adult mechanically ventilated patients
|
Clinical |
Circuit changes at 2-d intervals
|
Circuit changes at 30-d intervals
|
336 |
10.7 |
157 |
6.4 |
4 |
0.59 (0.30, 1.17) |
Han 200116 |
Adult mechanically ventilated patients
|
Clinical |
Circuit changes at 2-d intervals
|
Circuit changes at 7-d intervals
|
413 |
9.2 |
231 |
3.5 |
4 |
0.38 (0.18, 0.79) |
Lien 200117 |
Adult mechanically ventilated patients
|
Clinical |
Circuit changes at 2-d intervals
|
Circuit changes at 7-d intervals
|
6,213 |
2.9 |
7,068 |
3.2 |
4 |
1.14 (0.94, 1.38) |
TOTAL |
9,002 |
4.1 |
9,606 |
3.8 |
0.87 (0.63, 1.18) |
|||||
VAP = ventilator-associated
pneumonia. CI = confidence interval. |
||||||||||
Recommendation #2. Evidence is lacking related toVAP and issues of heated versus unheated circuits, type of heated humidifier, method for filling the humidifier, and technique for clearing condensate from the ventilator circuit. It is prudent to avoid excessive accumulation of condensate in the circuit. Care should be taken to avoid accidental drainage of condensate into the patient’s airway and to avoid contamination of caregivers during ventilator disconnection or during disposal of condensate. Care should be taken to avoid breaking the ventilator circuit, which could contaminate the interior of the circuit. (Grade D)
Is Ventilator-Associated Pneumonia Rate Affected By the Choice of Active Versus Passive Humidification in Mechanically Ventilated Patients?
Humidification of the inspired gas is a standard practice in the care of mechanically ventilated patients. Humidifiers can be active or passive. Active humidifiers pass the inspired gas either through (bubble) or over (passover, wick) a heated water bath. Passive humidifiers (artificial nose, heat-and-moisture exchanger) trap heat and humidity from the patient’s exhaled gas and return some of that to the patient on the subsequent inhalation. By its nature the ventilator circuit remains dry with the use of a passive humidifier. Passive humidifiers also have filtering characteristics, and some are constructed specifically to function as filters as well as humidifiers. The performance of humidification systems has been described in detail elsewhere.25
Because passive humidifiers maintain a dry circuit and have filtering properties, there has been much interest in their potential to decrease the incidence of VAP. Indeed, several studies have reported that circuit contamination is reduced with the use of passive humidifiers. 26 –31 One study reported similar tracheal colonization rates with active and passive humidifiers.32 The VAP rate with passive versus active humidification was addressed in 6 studies (Table 3).21,33–37 The combined results of these studies (Fig. 3) indicate a lower risk of VAP with the use of passive humidification (relative risk 0.65, 95% CI 0.44 to 0.96, p = 0.03). Although the combined effect from this meta-analysis shows a statistically lower VAP for the use of passive humidifiers, it is of interest to note that only one of the studies36 reported a significant reduction in VAP with the use of passive humidifiers. These studies used various brands of passive humidifier, all were conducted with adult patients, and all were conducted in the acute care setting. Two studies reported no significant difference in the VAP rate when comparing designs of passive humidifiers (different components, hydrophobic vs hygroscopic).38,39
In addition to VAP, there are other important issues that must be considered when a passive humidifier is used. These include the dead space of the device, the resistive load of the device, the difficulty in delivery of aerosolized medications, and the potential for airway occlusion. An increased work-of-breathing attributable to the use of a passive humidifier has been reported.40 The use of a passive humidifier has been associated with higher PaCO2 and higher minute ventilation requirement. 41 Another study reported significant reduction in PaCO2 with removal of the passive humidifier in patients receiving lung-protective ventilation.42 Of considerable concern is the increased risk of airway occlusion when a passive humidifier is used. In one study33 the use of passive humidifiers was interrupted because of a fatal occlusion of the airway. Other studies have also reported greater risk of airway occlusion with the use of a passive humidifier.34,43,44 A meta-analysis of airway occlusion associated with the use of passive humidifier reported a relative risk of 3.84 (95% CI 1.92 to 7.69, p = 0.0001), favoring the use of active humidification (ie, indicating a significantly greater risk of airway occlusion with a passive humidifier).45 When a passive humidifier is used, it is important that one be selected that has an adequate moisture output, to minimize the risk of airway occlusion.
Table 3. Summary of Randomized Controlled
Trials Investigating the
Relationship Between Type of Humidification and the Risk of Ventilator- Associated
Pneumonia
| Citation |
Study
Population |
VAP
Diagnosis |
Passive Humidifier |
Active Humidifier |
Passive Humidifier |
Level |
Relative
Risk (95% CI) |
||
| n |
Pneumonia
(%) |
n |
Pneumonia
(%) |
||||||
Martin 199033 |
Adult mechanically ventilated patients |
Clinical |
Pall Ultipor breathing circuit
filter |
42 |
19.0 |
31 |
6.5 |
2 |
0.34 (0.08, 1.49) |
Roustan 199234 |
Adult mechanically ventilated patients |
Clinical |
Pall BB 2215 |
61 |
14.8 |
55 |
9.1 |
1 |
0.62 (0.22, 1.73) |
Dreyfuss 199535 |
Adult mechanically ventilated patients |
Quantitative cultures |
DAR Hygrobac II |
70 |
11.4 |
61 |
9.8 |
1 |
0.86 (0.32, 2.34) |
Branson 199621 |
Adult mechanically ventilated patients |
Clinical |
Baxter nonfiltered hygroscopic
condenser humidifier |
54 |
5.6 |
49 |
6.1 |
2 |
1.10 (0.23, 5.21) |
Kirton 199736 |
Adult mechanically ventilated patients |
Clinical |
Pall BB-100 |
140 |
15.7 |
140 |
6.4 |
2 |
0.41 (0.20, 0.86) |
Kollef 199837 |
Adult mechanically ventilated patients |
Clinical |
Nellcor-Puritan-Bennett hygroscopic
condenser humidifier |
147 |
10.2 |
163 |
9.2 |
1 |
0.90 (0.46, 1.78) |
TOTAL |
514 |
12.6 |
499 |
8.0 |
0.65 (0.44, 0.96) |
||||
VAP = ventilator-associated
pneumonia. CI = confidence interval. |
|||||||||
Fig. 3. Randomized, controlled trials of the
relationship between type of
humidification and the risk of ventilator-associated pneumonia. RR = relative
risk. CI = confidence interval.

Recommendation #3. Although the available evidence suggests a lower VAP rate with passive humidification than with active humidification, other issues related to the use of passive humidifiers (resistance, dead space volume, airway occlusion risk) preclude a recommendation for the general use of these devices. The decision to use a passive humidifier should not be based solely on infection control considerations. (Grade A)
Do Passive Humidifiers Need to Be Changed at Regular Intervals?
The manufacturers of passive humidifiers typically recommend that they be changed at daily intervals. There has been interest in the safety of changing these devices less frequently, both in terms of VAP rate and device performance. Two randomized, controlled trials46,47 and 2 observational studies compared daily versus less frequent changes of passive humidifiers (Table 4).48,49 In 2 studies passive humidifiers were changed at 48-hour intervals,48,49 in a separate study they were changed at 5-day intervals,46 and in another study they were changed at 7-day intervals. 47 For the pooled results (Fig. 4), no significant difference in VAP rate was found with less frequent changes of passive humidifiers, in either the randomized, controlled studies (relative risk 0.58, 95% CI 0.24 to 1.41, p = 0.14) or the observational studies (relative risk 1.13, 95% CI 0.73 to 1.76, p = 0.9). Other studies have evaluated the technical performance of passive humidifiers used for durations up to 48 hours,50–52 96 hours,53–55 and 7 days.56 However, caution has been suggested related to prolonged use of passive humidifiers with some devices52,57 and in some patient populations (eg, chronic obstructive pulmonary disease).56 Based on the available evidence, it seems prudent to closely monitor the technical performance of these devices if used longer than 48 hours.
Recommendation #4. Passive humidifiers do not need to be changed daily for reasons of infection control or technical performance. They can be safely used for at least 48 hours, and with some patient populations some devices may be able to be used for up to 1 week. (Grade A)
Do In-Line Suction Catheters Need to Be Changed at Regular Intervals?
In-line closed suction systems allow mechanically ventilated patients to be suctioned without removal of ventilator support. This may decrease the complications associated with suctioning58 and might prevent alveolar derecruitment during suctioning.59,60 One study reported significantly less environmental contamination with closed suctioning than with open suctioning.61 Observational studies report high levels of contamination in closed suction catheters that are in use.62,63 However, this contamination usually arises from the endotracheal tube and the patient’s lower respiratory tract. Accordingly, the patient usually contaminates the catheter, rather than vice versa. Use of closed suctioning has been recommended as part of a VAPprevention program.64 Two prospective, randomized, controlled trials reported similar VAP rates with closed suctioning and open suctioning.58,65 Another study, however, reported a 3.5 times greater risk of VAP in patients randomized to receive open suctioning than those receiving closed suctioning.66 Although the available evidence is not conclusive that closed suctioning decreases the risk of VAP, there is no high-level evidence that use of closed suction catheters increases the risk of VAP.
The in-line closed suction catheter might be considered an extension of the ventilator circuit. Because ventilator circuits do not need to be changed at regular intervals for infection control purposes, this might suggest that in-line suction catheters also do not need to be changed at regular intervals for infection control purposes. Although the manufacturers of in-line suction catheters recommend that these devices be changed at regular intervals, there is accumulating evidence that they might not need to be changed routinely. One observational study reported no change in VAP rate when in-line suction catheters were changed on a weekly rather than daily basis.67 Another study reported no significant difference in VAP rate between patients randomized to receive daily changes of the in-line suction catheter and those with whom there were no routine changes of the in-line catheter (relative risk 0.99, 95% CI 0.66 to 1.50).68 The maximum duration of use of a closed suction catheter in that study was 67 days. There were few device malfunctions when in-line catheters were changed less frequently than daily, and there were important cost savings associated with this practice.67,68 As with ventilator circuits, the maximum duration that closed suction catheters can be used safely is not known.
Fig. 4. Studies of the relationship between
passive humidifier change frequency and the risk of
ventilator-associated pneumonia. RR = relative risk. CI = confidence interval.

Table 4. Summary of Studies Investigating
the Relationship Between Change Frequency
for Passive Humidifiers and the Risk of Ventilator- Associated Pneumonia
| Citation |
Study
Population |
VAP
Diagnosis |
Controld Group |
Treatment Group |
Control Group |
Treatment Group |
Level |
Relative
Risk (95% CI) |
||
| n |
Pneumonia
(%) |
n |
Pneumonia
(%) |
|||||||
Randomized Controlled
Trials
|
||||||||||
Davis 200046 |
Adult mechanically ventilated patients
|
Clinical |
HME changed every 24 h |
HME changed every 120 h |
100 |
8.0 |
120 |
7.5 |
1 |
0.94 (0.38, 2.34) |
Thomachot 200247 |
Adult mechanically ventilated patients
|
Clinical |
HME changed every 24 h |
HME changed every 7 d |
84 |
26.2 |
71 |
9.9 |
1 |
0.38 (0.17, 0.83) |
TOTAL |
184 |
16.3 |
191 |
8.4 |
0.58 (0.24, 1.41) |
|||||
Observational Studies
|
||||||||||
Djedaini 199548 |
Adult mechanically ventilated patients
|
Quantitative cultures |
HME changed every 24 h |
HME changed every 48 h |
61 |
9.8 |
68 |
11.8 |
4 |
1.20 (0.44, 3.25) |
Daumal 199949 |
Adult mechanically ventilated patients
|
Quantitative cultures |
HME changed every 24 h |
HME changed every 48 h |
174 |
14.4 |
187 |
16.0 |
4 |
1.12 (0.68, 1.82) |
TOTAL |
235 |
13.2 |
255 |
14.9 |
1.13 (0.73, 1.76) |
|||||
VAP = ventilator-associated
pneumonia. CI = confidence interval. |
||||||||||
Recommendation #5. The use of closed suction catheters should be considered part of a VAP prevention strategy. When closed suction catheters are used, they do not need to be changed daily for infection control purposes. The maximum duration of time that closed suction catheters can be used safely is unknown. (Grade A)
Other Issues
Other issues related to the technical aspects of mechanical ventilation may
be important in relation to VAP. Medication nebulizers can be a source of
contamination that could lead to VAP.69 Accordingly, with nebulizers
care must be taken to avoid contamination of the ventilator circuit and the
patient’s respiratory tract. It is commonly believed by respiratory
therapists that the risk of ventilator circuit contamination is reduced with
the use of metered-dose inhalers, but this has not been reported. Manual ventilator
devices that are commonly kept at the bedside of mechanically ventilated patients
have been shown to be a source of airway contamination, so care should be
taken to minimize the potential infection risks associated with manual ventilator
devices. 70,71 One study reported a significantly greater likelihood
of VAP among patients who underwent transport out of the intensive care unit
for diagnostic, surgical, or miscellaneous interventions (odds ratio 3.8,
95% CI 2.6 to 5.5, p = 0.001).72 Because a VAP risk education program
for respiratory therapists and critical care nurses decreased the incidence
of VAP,73 VAP risk education should be widely implemented.
Recommendation # 6. Clinicians (respiratory therapists, nurses, and physicians) caring for mechanically ventilated patients should be aware of risk factors for VAP (eg, nebulizer therapy, manual ventilation, and patient transport). (Grade B)
Discussion
Substantial evidence now exists to make recommendations related to
the technical practices of mechanical ventilation and the risk of VAP. However,
important gaps also exist in this evidence base. For example, most of the
published evidence comes from studies of adult patients. Few data have been
published for neonatal and pediatric populations. In addition most studies
come from the acute care setting. Finally, there has been little research
done on important subgroups of patients, such as immunocompromised patients.
Accordingly, much remains to be learned about important relationships between
the technical aspects of mechanical ventilation and the risk of VAP. Thus,
these evidence-based guidelines provide not only a basis for current practice
but are also a framework for further investigation.
REFERENCES
Interested persons may copy these Guidelines for noncommercial purposes of scientific or educational advancement. Please credit AARC and Respiratory Care Journal.
Reprinted from the September 2003 issue of RESPIRATORY CARE [Respir Care 2003;48(9):869–879]
You are here: RCJournal.com » Clinical Practice Guidelines